Medline Industries, LP Recall: Medline Kits containing Tego Connectors Medline DIALYSIS ON is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Medline Industries, LP; Product: Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746; Reason for recall: Affected lots
| Recalling firm | Medline Industries, LP |
|---|---|
| Product | Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746 |
| Reason for recall | Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | US Nationwide distribution. |
| Recall initiated | 20260108 |
| Recall number | Z-1464-2026 |
| Category | device |
| State | IL |
| Country | United States |