Medtronic Navigation, Inc. Recall: StealthStation Cranial and Synergy Cranial Depth Gauge:

Medtronic Navigation, Inc. Recall: StealthStation Cranial and Synergy Cranial Depth Gauge: is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Medtronic Navigation, Inc.; Product: StealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide - Instruct

Recalling firmMedtronic Navigation, Inc.
ProductStealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide - Instructions for cranial procedures using optical localization StealthStation Cranial version 3.0, 3.1 CE 0344 Rx Only 2018-02 9735489 Revision 3 Medtronic Navigation, Inc. Biopsy Needle Kit (9733068) Note: This device is an accessory to your StealthStation System. . . The Biopsy Needle is a presterilized , single-use, side-cutting needle for cranial tissue biopsies. CE 0344 Rx Only 2017-05 9733068-01 Medtronic Navigation, Inc.
Reason for recallAn incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.
ClassificationClass I (most serious — reasonable probability of serious health consequences or death)
StatusTerminated
DistributionUS: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY OUS: Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Bar
Recall initiated20180926
Recall numberZ-0552-2019
Categorydevice
StateCO
CountryUnited States