Medtronic Neuromodulation Recall: Synchromed II programmable pump, Model numbers 8637-20 a
Medtronic Neuromodulation Recall: Synchromed II programmable pump, Model numbers 8637-20 a is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Medtronic Neuromodulation; Product: Synchromed II programmable pump, Model numbers 8637-20 and 8637-40; Reason for recall: There is a potential for
| Recalling firm | Medtronic Neuromodulation |
|---|---|
| Product | Synchromed II programmable pump, Model numbers 8637-20 and 8637-40 |
| Reason for recall | There is a potential for the presence of a foreign particle inside the pump motor assembly which could interfere with the motor gear rotation and lead to a permanent motor stall. |
| Classification | Class I (most serious — reasonable probability of serious health consequences or death) |
| Status | Terminated |
| Distribution | US and Canada, Asia -Pacific (APAC), China, Europe, Middle East and Africa (EMAE), Latin America (LATAM) |
| Recall initiated | 20191011 |
| Recall number | Z-0508-2020 |
| Category | device |
| State | MN |
| Country | United States |