Nico Corp. Recall: Myriad-LX Light Source-for delivery of light to the surgical field to

Nico Corp. Recall: Myriad-LX Light Source-for delivery of light to the surgical field to is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Nico Corp.; Product: Myriad-LX Light Source-for delivery of light to the surgical field to enhance visualization of tissue during resection Part Numb

Recalling firmNico Corp.
ProductMyriad-LX Light Source-for delivery of light to the surgical field to enhance visualization of tissue during resection Part Number NN-7013 The Myriad-LX light source is a high-intensity 300W Xenon illuminator with a touch screen and digital dimming controls. The source is fitted with a proprietary connector to ensure usage is reserved for the Myriad-LX Illumination Fiber.
Reason for recall(1) Updated IFU for the LX Illumination Packs: Warning - Avoid contact between the distal opening of the delivery sleeve and tissue/debris within the surgical.field to prevent complete occlusion. Some surgical approaches (e.g., endonasal) may increase this potential. {f unintended reduction in light output occurs and/or occlusion of the sleeve is suspected, both the delivery sleeve and the fiber should be replaced (2) Removal for the LX Light Source potential for the distal tip of the sleeve/fiber assembly potential to burn healthy tissue resulting in minor to moderate injury to the patient or
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionUS Nationwide Distribution
Recall initiated20210430
Recall numberZ-1840-2021
Categorydevice