Nuvasive Specialized Orthopedics Inc Recall: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. Precic
Nuvasive Specialized Orthopedics Inc Recall: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. Precic is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Nuvasive Specialized Orthopedics Inc; Product: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. Precice Bone Transport Antegrade Femur Trochanter 10 Degree B
| Recalling firm | Nuvasive Specialized Orthopedics Inc |
|---|---|
| Product | NUVASIVE SPECIALIZED ORTHOPEDICS, INC. Precice Bone Transport Antegrade Femur Trochanter 10 Degree Bend Component of the orthopedic Precice Bone Transport System of intramedullary nail, locking screws, reusable instruments, and hand-held External Remote Controller (ERC). |
| Reason for recall | During explantation, standard removal tools can generate sufficient forces to cause distal plug to disassociate from the bone transport system. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide distribution. US nationwide, Austria, Australia, Canada, Switzerland, Denmark, Spain, France, United Kingdom, Germany, UAE, Luxembourg, Israel, Netherlands. |
| Recall initiated | 20200724 |
| Recall number | Z-0069-2021 |
| Category | device |
| State | CA |
| Country | United States |