Olympus Scientific Solutions Americas Recall: Olympus Scientific Solutions Americas Corpor
Olympus Scientific Solutions Americas Recall: Olympus Scientific Solutions Americas Corpor is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Olympus Scientific Solutions Americas; Product: Olympus Scientific Solutions Americas Corporation ( OSSA ) VANTA¿ XRF Analyzer. Model Vanta VCR and
| Recalling firm | Olympus Scientific Solutions Americas |
|---|---|
| Product | Olympus Scientific Solutions Americas Corporation ( OSSA ) VANTA¿ XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems |
| Reason for recall | It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | US Distribution to the states of : PA, MA, CO, OH, NJ, GA, CA, TX, IL, VA, MN, NH, WA, SC, AZ, LA, MD, OK, IN, NY, NC and KS. |
| Recall initiated | 20170107 |
| Recall number | Z-1036-2017 |
| Category | device |
| State | MA |
| Country | United States |
$29/mo