← RecallGuard
HomeRecalls

Olympus Scientific Solutions Americas Recall: Olympus Scientific Solutions Americas Corpor

Olympus Scientific Solutions Americas Recall: Olympus Scientific Solutions Americas Corpor is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Olympus Scientific Solutions Americas; Product: Olympus Scientific Solutions Americas Corporation ( OSSA ) VANTA¿ XRF Analyzer. Model Vanta VCR and

Recalling firmOlympus Scientific Solutions Americas
ProductOlympus Scientific Solutions Americas Corporation ( OSSA ) VANTA¿ XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems
Reason for recallIt was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionUS Distribution to the states of : PA, MA, CO, OH, NJ, GA, CA, TX, IL, VA, MN, NH, WA, SC, AZ, LA, MD, OK, IN, NY, NC and KS.
Recall initiated20170107
Recall numberZ-1036-2017
Categorydevice
StateMA
CountryUnited States

🔍 Search all recalls →