Orthalign, Inc Recall: Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 c
Orthalign, Inc Recall: Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 c is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Orthalign, Inc; Product: Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System; Reason for recall: Drill
| Recalling firm | Orthalign, Inc |
|---|---|
| Product | Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System |
| Reason for recall | Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm) |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | US Nationwide distribution in the state of UT. |
| Recall initiated | 20260210 |
| Recall number | Z-1630-2026 |
| Category | device |
| State | CA |
| Country | United States |