Phadia US Inc Recall: Phadia 1000 Instrument, Article Number 12-3800-01.
Phadia US Inc Recall: Phadia 1000 Instrument, Article Number 12-3800-01. is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Phadia US Inc; Product: Phadia 1000 Instrument, Article Number 12-3800-01.; Reason for recall: The "Retry" command does not function properly which could cause a sho
| Recalling firm | Phadia US Inc |
|---|---|
| Product | Phadia 1000 Instrument, Article Number 12-3800-01. |
| Reason for recall | The "Retry" command does not function properly which could cause a shortage of Wash and Rinse solution and affect assay performance and test results. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Nationwide |
| Recall initiated | 20170705 |
| Recall number | Z-0387-2018 |
| Category | device |
| State | MI |
| Country | United States |