Philips Medical Systems (Cleveland) Inc Recall: BrightView X 3/4 model number 882479 Brigh
Philips Medical Systems (Cleveland) Inc Recall: BrightView X 3/4 model number 882479 Brigh is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Philips Medical Systems (Cleveland) Inc; Product: BrightView X 3/4 model number 882479 BrightView and BrightView X are gamma cameras designed for s
| Recalling firm | Philips Medical Systems (Cleveland) Inc |
|---|---|
| Product | BrightView X 3/4 model number 882479 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only). BrightView XCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates with |
| Reason for recall | The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide distribution - US including AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NY, OH, OR, PA, TN, TX, VA, WA, WI, and WV. Internationally to Australia, Austria, Belgium, Brazil, Canada, Chile, China, Denmark, Egypt, France, Germany, India, Israel, Italy, Japan, Me |
| Recall initiated | 20170224 |
| Recall number | Z-1752-2017 |
| Category | device |
| State | OH |
| Country | United States |