Philips Medical Systems (Cleveland) Inc Recall: Ingenuity Core 128, (Code 728323) Intended

Philips Medical Systems (Cleveland) Inc Recall: Ingenuity Core 128, (Code 728323) Intended is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Philips Medical Systems (Cleveland) Inc; Product: Ingenuity Core 128, (Code 728323) Intended to produce images of the head and body by computer rec

Recalling firmPhilips Medical Systems (Cleveland) Inc
ProductIngenuity Core 128, (Code 728323) Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason for recallPhilips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution -- US, Argentina, China, Republic of Korea, Russian Federation, Slovakia, Turkey, and United Kingdom. 4 in US, (1 military), 14 foreign accounts
Recall initiated20171110
Recall numberZ-0391-2018
Categorydevice
StateOH
CountryUnited States