PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Recall: Intera 1.5T Master/Nova, Model Number: 7811

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Recall: Intera 1.5T Master/Nova, Model Number: 7811 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Product: Intera 1.5T Master/Nova, Model Number: 781106;; Reason for recall: The Quadrature Body Coil (QBC)

Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
ProductIntera 1.5T Master/Nova, Model Number: 781106;
Reason for recallThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS: Nationwide Distribution International: Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Re
Recall initiated20250324
Recall numberZ-1776-2025
Categorydevice
CountryNetherlands