Philips North America Llc Recall: Brilliance iCT; Product Code (REF): 728306;
Philips North America Llc Recall: Brilliance iCT; Product Code (REF): 728306; is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Philips North America Llc; Product: Brilliance iCT; Product Code (REF): 728306;; Reason for recall: As part of the CT systems sub assembly, thirty-two (32) fast
| Recalling firm | Philips North America Llc |
|---|---|
| Product | Brilliance iCT; Product Code (REF): 728306; |
| Reason for recall | As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egy |
| Recall initiated | 20260107 |
| Recall number | Z-1315-2026 |
| Category | device |
| State | MA |
| Country | United States |