Philips North America, LLC Recall: PageWriter TC70, Product 860315, Software revisions up
Philips North America, LLC Recall: PageWriter TC70, Product 860315, Software revisions up is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Philips North America, LLC; Product: PageWriter TC70, Product 860315, Software revisions up to and including A.07.05.22 used to evaluate the electro
| Recalling firm | Philips North America, LLC |
|---|---|
| Product | PageWriter TC70, Product 860315, Software revisions up to and including A.07.05.22 used to evaluate the electrocardiogram of adult and pediatric patients |
| Reason for recall | Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Domestic distribution nationwide. Foreign distribution worldwide. |
| Recall initiated | 20190109 |
| Recall number | Z-0853-2019 |
| Category | device |
| State | MA |
| Country | United States |