Philips North America Recall: Intera 1.5T Achieva Nova. Model (REF) Numbers 781172. is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Philips North America; Product: Intera 1.5T Achieva Nova. Model (REF) Numbers 781172.; Reason for recall: The Quadrature Body Coil (QBC) seal adhesive may
| Recalling firm | Philips North America |
|---|---|
| Product | Intera 1.5T Achieva Nova. Model (REF) Numbers 781172. |
| Reason for recall | The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide - US Nationwide distribution. |
| Recall initiated | 20240103 |
| Recall number | Z-0952-2024 |
| Category | device |
| State | MA |
| Country | United States |