Philips North America Recall: Spectral CT on Rails, Software Version 5.1.0, Model Number 7 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Philips North America; Product: Spectral CT on Rails, Software Version 5.1.0, Model Number 728334; Reason for recall: This recall is part of a 2-is
| Recalling firm | Philips North America |
|---|---|
| Product | Spectral CT on Rails, Software Version 5.1.0, Model Number 728334 |
| Reason for recall | This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID93363 |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | US Nationwide distribution in the state of MN. |
| Recall initiated | 20231009 |
| Recall number | Z-0319-2024 |
| Category | device |
| State | MA |
| Country | United States |