Philips Respironics, Inc. Recall: Trilogy Evo Universal, Product number DS2000X11B

Philips Respironics, Inc. Recall: Trilogy Evo Universal, Product number DS2000X11B is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Philips Respironics, Inc.; Product: Trilogy Evo Universal, Product number DS2000X11B; Reason for recall: Two software issues have been identified related t

Recalling firmPhilips Respironics, Inc.
ProductTrilogy Evo Universal, Product number DS2000X11B
Reason for recallTwo software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionWorldwide distribution - US nationwide distribution.
Recall initiated20210726
Recall numberZ-2290-2021
Categorydevice
StatePA
CountryUnited States