Pointe Scientific, Inc. Recall: Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5

Pointe Scientific, Inc. Recall: Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Pointe Scientific, Inc.; Product: Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Als

Recalling firmPointe Scientific, Inc.
ProductPointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass vials Cap: Rubber seal with colored aluminum seals Glucose-6-phosphate dehydrogenase (G6PD) controls can be used to test for the quantitative and qualitative determination of G6PD in blood.
Reason for recallA reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current reconstitution claim of 7 days. For this reason we suggest discontinuing use of the control set.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide distribution: US (nationwide) including states of: PA, CA, KY, FL, NJ, DE, MI, MN,and IL; and countries of: United Arab Emirates, Austria, Italy, Israel, Greece, and Australia.
Recall initiated20150519
Recall numberZ-2487-2015
Categorydevice
StateMI
CountryUnited States