Qiagen Sciences LLC Recall: therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121

Qiagen Sciences LLC Recall: therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Qiagen Sciences LLC; Product: therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121; Reason for recall: There is a risk for a false mutati

Recalling firmQiagen Sciences LLC
Producttherascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121
Reason for recallThere is a risk for a false mutation positive result in rare cases resulting from a fluorescence artefact. In these rare cases, the controls pass correctly, while a fluorescence artefact in the mutation assay causes an incorrect valid mutation positive result.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionUS Nationwide.
Recall initiated20200408
Recall numberZ-2290-2020
Categorydevice
StateMD
CountryUnited States