Qiagen Sciences LLC Recall: therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121
Qiagen Sciences LLC Recall: therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Qiagen Sciences LLC; Product: therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121; Reason for recall: There is a risk for a false mutati
| Recalling firm | Qiagen Sciences LLC |
|---|---|
| Product | therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121 |
| Reason for recall | There is a risk for a false mutation positive result in rare cases resulting from a fluorescence artefact. In these rare cases, the controls pass correctly, while a fluorescence artefact in the mutation assay causes an incorrect valid mutation positive result. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | US Nationwide. |
| Recall initiated | 20200408 |
| Recall number | Z-2290-2020 |
| Category | device |
| State | MD |
| Country | United States |