← RecallGuard
HomeRecalls

Randox Laboratories Ltd. Recall: Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro d

Randox Laboratories Ltd. Recall: Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro d is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Randox Laboratories Ltd.; Product: Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Comp

Recalling firmRandox Laboratories Ltd.
ProductRandox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgM, Prealbumin, RF and Transferrin assays Catalogue Number: IT2691
Reason for recallAs part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionCA, FL, IL, ME, MI, NC, OH, PA,NJ,
Recall initiated20230719
Recall numberZ-2448-2023
Categorydevice
StateN/A
CountryIreland

🔍 Search all recalls →