← RecallGuard
HomeRecalls

RAYSEARCH LABORATORIES AB Recall: RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1

RAYSEARCH LABORATORIES AB Recall: RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: RAYSEARCH LABORATORIES AB; Product: RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system; Re

Recalling firmRAYSEARCH LABORATORIES AB
ProductRayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system
Reason for recallThere is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionThe products were distributed to the following US states: TN
Recall initiated20200220
Recall numberZ-1463-2020
Categorydevice
CountrySweden