RAYSEARCH LABORATORIES AB Recall: RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1
RAYSEARCH LABORATORIES AB Recall: RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: RAYSEARCH LABORATORIES AB; Product: RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system; Re
| Recalling firm | RAYSEARCH LABORATORIES AB |
|---|---|
| Product | RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system |
| Reason for recall | There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | The products were distributed to the following US states: TN |
| Recall initiated | 20200220 |
| Recall number | Z-1463-2020 |
| Category | device |
| Country | Sweden |