RAYSEARCH LABORATORIES AB Recall: RayStation Therapy Treatment Planning System, Model Numb

RAYSEARCH LABORATORIES AB Recall: RayStation Therapy Treatment Planning System, Model Numb is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: RAYSEARCH LABORATORIES AB; Product: RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Pla

Recalling firmRAYSEARCH LABORATORIES AB
ProductRayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. RayStation is a software system designed for treatment planning and analysis of radiation therapy.
Reason for recallUnder some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avoid incorrect dose calculations during treatment planning.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution - USA including Connecticut, Florida, Massachusetts, Missouri, New York, Ohio, and foreign distribution to(Canada, Italy, Australia, Germany, Netherlands, and Switzerland.
Recall initiated20130215
Recall numberZ-1419-2013
Categorydevice
StateN/A
CountrySweden