SEDECAL SA Recall: MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: SEDECAL SA; Product: MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007; Reason for recall: While cleaning or disinfecting

Recalling firmSEDECAL SA
ProductMobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007
Reason for recallWhile cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS Nationwide distribution. There was government distribution but no military distribution.
Recall initiated20211014
Recall numberZ-1083-2024
Categorydevice