SenTec AG Recall: Sentec, V-Sign Sensor 2 (REF VS-A/P/N), Reusable / Nonsterile / Waterp

SenTec AG Recall: Sentec, V-Sign Sensor 2 (REF VS-A/P/N), Reusable / Nonsterile / Waterp is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: SenTec AG; Product: Sentec, V-Sign Sensor 2 (REF VS-A/P/N), Reusable / Nonsterile / Waterproof, SenTec AG, CH-4106, Therwil, Switzerland. The SenTec

Recalling firmSenTec AG
ProductSentec, V-Sign Sensor 2 (REF VS-A/P/N), Reusable / Nonsterile / Waterproof, SenTec AG, CH-4106, Therwil, Switzerland. The SenTec Digital Monitoring System (SDMS) is used for continuous and non-invasive monitoring of transcutaneous carbon dioxide tension (tcPCO2), oxygen saturation (SpO2), and pulse rate (PR).
Reason for recallSenTec is voluntarily replacing V-Sign Sensors 2 due to pre-mature deterioration of a part of the PCO2 measuring unit and thus disabling PCO2-monitoring. This means that the affected sensors reach their end-of-life earlier than normally expected requiring the sensor`s replacement. There is no increased risk for the patient or operator, because the aforementioned deterioration results in the sens
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionUS Distribution to the state of : MO
Recall initiated20150325
Recall numberZ-1501-2015
Categorydevice
CountrySwitzerland