Sentinel CH SpA Recall: Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 0

Sentinel CH SpA Recall: Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 0 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Sentinel CH SpA; Product: Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 07P5605) - Product Usage: use in the calibration of the

Recalling firmSentinel CH SpA
ProductAlinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 07P5605) - Product Usage: use in the calibration of the Alinity c CRP Vario assay, Cardiac CRP high sensitivity method [Cardiac CRP], on the Alinity c analyzer.
Reason for recallCalibrator vials were incorrectly labeled. Use of incorrectly labeled vials will cause the instrument to generate a warning that calibrators are incorrectly loaded and calibration will not be completed successfully.
ClassificationClass III (least serious — unlikely to cause adverse health consequences)
StatusTerminated
DistributionUS Nationwide distribution in the states of LA, MD, MN, ND, NY, WV, and Puerto Rico.
Recall initiated20210729
Recall numberZ-0033-2022
Categorydevice
CountryItaly