← RecallGuard
HomeRecalls

Shimadzu Corp. Analytical & Measuring Instruments Recall: Autosampler, Models: SIL-40C XR

Shimadzu Corp. Analytical & Measuring Instruments Recall: Autosampler, Models: SIL-40C XR is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Shimadzu Corp. Analytical & Measuring Instruments Div.; Product: Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL; Reason for recall: There have be

Recalling firmShimadzu Corp. Analytical & Measuring Instruments Div.
ProductAutosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
Reason for recallThere have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionWorldwide - US Nationwide distribution in the states of DE, PA and the country of DE.
Recall initiated20241009
Recall numberZ-0554-2025
Categorydevice
CountryJapan

🔍 Search all recalls →