Siemens Healthcare Diagnostics, Inc. Recall: ADVIA Chemistry Urinary/Cerebrospinal Fluid P
Siemens Healthcare Diagnostics, Inc. Recall: ADVIA Chemistry Urinary/Cerebrospinal Fluid P is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Siemens Healthcare Diagnostics, Inc.; Product: ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagn
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
|---|---|
| Product | ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the quantitative determination of total protein in human urine and cerebrospinal fluid |
| Reason for recall | There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed after the UCFP test on ADVIA Chemistry systems. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide - US Nationwide distribution in the states of CA, CO, CT, FL, GA, IA, MD, MI, NJ, NM, NY, OH, PA, SD, TX, UT, and VA. Global distribution. |
| Recall initiated | 20230725 |
| Recall number | Z-2451-2023 |
| Category | device |
| State | NY |
| Country | United States |