Siemens Healthcare Diagnostics, Inc. Recall: ADVIA Chemistry Urinary/Cerebrospinal Fluid P

Siemens Healthcare Diagnostics, Inc. Recall: ADVIA Chemistry Urinary/Cerebrospinal Fluid P is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Siemens Healthcare Diagnostics, Inc.; Product: ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagn

Recalling firmSiemens Healthcare Diagnostics, Inc.
ProductADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the quantitative determination of total protein in human urine and cerebrospinal fluid
Reason for recallThere is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed after the UCFP test on ADVIA Chemistry systems.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionWorldwide - US Nationwide distribution in the states of CA, CO, CT, FL, GA, IA, MD, MI, NJ, NM, NY, OH, PA, SD, TX, UT, and VA. Global distribution.
Recall initiated20230725
Recall numberZ-2451-2023
Categorydevice
StateNY
CountryUnited States