Siemens Healthcare Diagnostics, Inc. Recall: Magnesium Flex reagent cartridge, 360 tests,
Siemens Healthcare Diagnostics, Inc. Recall: Magnesium Flex reagent cartridge, 360 tests, is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Siemens Healthcare Diagnostics, Inc.; Product: Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
|---|---|
| Product | Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158; |
| Reason for recall | Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Slovenia, Spain. |
| Recall initiated | 20241112 |
| Recall number | Z-0767-2025 |
| Category | device |
| State | DE |
| Country | United States |