Siemens Healthcare Diagnostics Inc Recall: Siemens RAPIDPoint 500 Measurement Cartridge La

Siemens Healthcare Diagnostics Inc Recall: Siemens RAPIDPoint 500 Measurement Cartridge La is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Siemens Healthcare Diagnostics Inc; Product: Siemens RAPIDPoint 500 Measurement Cartridge Lactate (750 tests) Part Number: 10491449 Product Usage:

Recalling firmSiemens Healthcare Diagnostics Inc
ProductSiemens RAPIDPoint 500 Measurement Cartridge Lactate (750 tests) Part Number: 10491449 Product Usage: The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCHb, FMetHb, FHHb, nBili
Reason for recallRAPIDPoint 500 measurement cartridge serial numbers 2517102517 through 2519000012, negative sodium bias on both QC and patient results
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution - US Nationwide and the countries of Canada, Albania, Austria, Australia, Bosnia Herzeg., Bangladesh, Belgium, Brazil, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Denmark, Estonia, Spain, Finland, France, Great Britain, Georgia, Greece, Croatia, Hung
Recall initiated20150824
Recall numberZ-2813-2015
Categorydevice
StateMA
CountryUnited States