Siemens Medical Solutions USA, Inc Recall: Artis Systems with Software Artis VC20x/VC21A/V is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Siemens Medical Solutions USA, Inc; Product: Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-
| Recalling firm | Siemens Medical Solutions USA, Inc |
|---|---|
| Product | Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system |
| Reason for recall | The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Nationwide Distribution including AL, CA, CO, FL, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, ND, NJ, NY, OH, PA, SC, TX, UT, VA, WA, and WI and to Puerto Rico. |
| Recall initiated | 20130308 |
| Recall number | Z-0997-2013 |
| Category | device |
| State | PA |
| Country | United States |