Siemens Medical Solutions USA, Inc Recall: Artis Systems with Software Artis VC20x/VC21A/V

Siemens Medical Solutions USA, Inc Recall: Artis Systems with Software Artis VC20x/VC21A/V is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Siemens Medical Solutions USA, Inc; Product: Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-

Recalling firmSiemens Medical Solutions USA, Inc
ProductArtis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system
Reason for recallThe firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionNationwide Distribution including AL, CA, CO, FL, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, ND, NJ, NY, OH, PA, SC, TX, UT, VA, WA, and WI and to Puerto Rico.
Recall initiated20130308
Recall numberZ-0997-2013
Categorydevice