← RecallGuard
HomeRecalls

Siemens Medical Solutions USA, Inc Recall: Biograph mMR. Model Number: 10433372.

Siemens Medical Solutions USA, Inc Recall: Biograph mMR. Model Number: 10433372. is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Siemens Medical Solutions USA, Inc; Product: Biograph mMR. Model Number: 10433372.; Reason for recall: There is a potential for an ice blockage to form or cu

Recalling firmSiemens Medical Solutions USA, Inc
ProductBiograph mMR. Model Number: 10433372.
Reason for recallThere is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
ClassificationClass I (most serious — reasonable probability of serious health consequences or death)
StatusOngoing
DistributionWorldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyp
Recall initiated20250828
Recall numberZ-2654-2025
Categorydevice
StatePA
CountryUnited States

🔍 Search all recalls →