Siemens Medical Solutions USA, Inc Recall: Luminos dRF Max with software VE10, VF10, VF11-
Siemens Medical Solutions USA, Inc Recall: Luminos dRF Max with software VE10, VF10, VF11- is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Siemens Medical Solutions USA, Inc; Product: Luminos dRF Max with software VE10, VF10, VF11- A diagnostic imaging system for radiographic and fluor
| Recalling firm | Siemens Medical Solutions USA, Inc |
|---|---|
| Product | Luminos dRF Max with software VE10, VF10, VF11- A diagnostic imaging system for radiographic and fluoroscopic studie Model Number:10762471 |
| Reason for recall | Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | US Nationwide distribution. |
| Recall initiated | 20221221 |
| Recall number | Z-1167-2023 |
| Category | device |
| State | PA |
| Country | United States |