Siemens Medical Solutions USA, Inc Recall: Siemens Artis zee Ceiling Systems. Angiographic
Siemens Medical Solutions USA, Inc Recall: Siemens Artis zee Ceiling Systems. Angiographic is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Siemens Medical Solutions USA, Inc; Product: Siemens Artis zee Ceiling Systems. Angiographic x-ray system.; Reason for recall: During the lifetime
| Recalling firm | Siemens Medical Solutions USA, Inc |
|---|---|
| Product | Siemens Artis zee Ceiling Systems. Angiographic x-ray system. |
| Reason for recall | During the lifetime of Artis zee Ceiling systems, there is an increased wear due to potential sub-optimal cable guidance that can lead to cable damage. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Nationwide Distribution. |
| Recall initiated | 20131216 |
| Recall number | Z-1080-2014 |
| Category | device |
| State | PA |
| Country | United States |