Siemens Medical Solutions USA, Inc. Recall: The following Pro.specta Emission Computed Tom

Siemens Medical Solutions USA, Inc. Recall: The following Pro.specta Emission Computed Tom is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Siemens Medical Solutions USA, Inc.; Product: The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look

Recalling firmSiemens Medical Solutions USA, Inc.
ProductThe following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical Sensor Assembly: (a) Symbia Pro.specta Q3, Model 11364751; (b) Symbia Pro.specta X3, Model 11364752; (c) Symbia Pro.specta X7, Model 11364753.
Reason for recallDuring a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the patient than needed leading to a reduction in image resolution.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionDistribution was made to ID, KS, MA, NH, OH, TX, and VA. There was no government/military distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Canada, Colombia, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, Netherlands, Portugal, Singapore, Swi
Recall initiated20230712
Recall numberZ-2256-2023
Categorydevice