← RecallGuard
HomeRecalls

Smith & Nephew, Inc. Recall: smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Sm

Smith & Nephew, Inc. Recall: smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Sm is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Smith & Nephew, Inc.; Product: smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2

Recalling firmSmith & Nephew, Inc.
Productsmith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STOUT STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442030, QTY: (1), STERILE R Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (VLP MINI-MOD) are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.
Reason for recallDevices, incorrectly labeled as a VLP MINI-MOD 2.0MM STRAIGHT PLATE 8 HOLE, contained a VLP MINI-MOD 2.0MM STOUT PLATE 8 HOLE and vice versa
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionUS Distribution to states of: MI, GA, and OH.
Recall initiated20171211
Recall numberZ-0397-2018
Categorydevice
StateTN
CountryUnited States

🔍 Search all recalls →