St. Jude Medical, Inc. Recall: Infinity 7, Implantable Pulse Generator, REF 6663, Rx only,

St. Jude Medical, Inc. Recall: Infinity 7, Implantable Pulse Generator, REF 6663, Rx only, is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: St. Jude Medical, Inc.; Product: Infinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an elect

Recalling firmSt. Jude Medical, Inc.
ProductInfinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommod
Reason for recallThe firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionUS Nationwide Distribution
Recall initiated20170602
Recall numberZ-2696-2017
Categorydevice
StateTX
CountryUnited States