Stryker Corporation Recall: Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Stryker Corporation; Product: Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for general electrosurgical ap
| Recalling firm | Stryker Corporation |
|---|---|
| Product | Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-000 |
| Reason for recall | There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Ongoing |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia. |
| Recall initiated | 20241004 |
| Recall number | Z-0365-2025 |
| Category | device |
| State | MI |
| Country | United States |