Stryker Endoscopy Recall: Stryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Stryker Endoscopy; Product: Stryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning a

Recalling firmStryker Endoscopy
ProductStryker Irrigator Reusable Tips: 5mm X 24cm Regular Tip, Vented, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
Reason for recallThe pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.
Recall initiated20130327
Recall numberZ-1224-2013
Categorydevice
StateCA
CountryUnited States