Stryker Neurovascular Recall: Target Detachable Coils: TARGET 360 NANO 1.5 MM X 4 CM MODEL
Stryker Neurovascular Recall: Target Detachable Coils: TARGET 360 NANO 1.5 MM X 4 CM MODEL is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Stryker Neurovascular; Product: Target Detachable Coils: TARGET 360 NANO 1.5 MM X 4 CM MODEL Number:M0035421540 Neurology: Target Detachable Coils
| Recalling firm | Stryker Neurovascular |
|---|---|
| Product | Target Detachable Coils: TARGET 360 NANO 1.5 MM X 4 CM MODEL Number:M0035421540 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. |
| Reason for recall | Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahi |
| Recall initiated | 20160606 |
| Recall number | Z-2164-2016 |
| Category | device |
| State | CA |
| Country | United States |
$29/mo