← RecallGuard
HomeRecalls

Sysmex America, Inc. Recall: Sysmex CV-11 Sample Unit, an integrated modular sampler that

Sysmex America, Inc. Recall: Sysmex CV-11 Sample Unit, an integrated modular sampler that is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Sysmex America, Inc.; Product: Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyze

Recalling firmSysmex America, Inc.
ProductSysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.
Reason for recallWhen excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionUSA: FL, VT, IL.
Recall initiated20200121
Recall numberZ-1717-2020
Categorydevice
StateIL
CountryUnited States

🔍 Search all recalls →