← RecallGuard
HomeRecalls

TECHNO-PATH MANUFACTURING LTD. Recall: Multichem P, Part Code 08P90-10

TECHNO-PATH MANUFACTURING LTD. Recall: Multichem P, Part Code 08P90-10 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: TECHNO-PATH MANUFACTURING LTD.; Product: Multichem P, Part Code 08P90-10; Reason for recall: Technopath Manufacturing Ltd. has identified an issue with the glass vial

Recalling firmTECHNO-PATH MANUFACTURING LTD.
ProductMultichem P, Part Code 08P90-10
Reason for recallTechnopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusOngoing
DistributionUS Nationwide distribution in the state of Illinois.
Recall initiated20230825
Recall numberZ-0106-2024
Categorydevice
StateN/A
CountryIreland

🔍 Search all recalls →