Teleflex Medical Recall: Hudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-
Teleflex Medical Recall: Hudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5- is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Teleflex Medical; Product: Hudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REF 5-10313 e)
| Recalling firm | Teleflex Medical |
|---|---|
| Product | Hudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REF 5-10313 e) 7.0 mm, REF 5-10314 f) 7.5 mm, REF 5-10315 g) 8.0 mm, REF 5-10316 h) 8.5 mm, REF 5-10317 i) 9.0 mm, REF 5-10318 j) 10.0 mm, REF 5-10320 Product Usage: Tracheal tube/airway management |
| Reason for recall | Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention. |
| Classification | Class I (most serious — reasonable probability of serious health consequences or death) |
| Status | Terminated |
| Distribution | US Nationwide Distribution |
| Recall initiated | 20190730 |
| Recall number | Z-2316-2019 |
| Category | device |
| State | NC |
| Country | United States |