Terumo Cardiovascular Systems Corporation Recall: Battery, 220V (Includes charger and conn
Terumo Cardiovascular Systems Corporation Recall: Battery, 220V (Includes charger and conn is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Terumo Cardiovascular Systems Corporation; Product: Battery, 220V (Includes charger and connecting cable) The Sarns" Centrifugal System is indicate
| Recalling firm | Terumo Cardiovascular Systems Corporation |
|---|---|
| Product | Battery, 220V (Includes charger and connecting cable) The Sarns" Centrifugal System is indicated for use in cardiopulmonary bypass procedures only. |
| Reason for recall | Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2004 of various malfunctions of the battery backup system for the Sarns" Centrifugal System. The reports can be broadly categorized as follows: " Failure to charge " Failure to hold charge " Failure to operate In the most severe case, a battery malfunction could result in an unexpected loss of arterial flow. In |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide distribution: USA including: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, UT, VA, WA, WI, and WV. Puerto Rico and internationally to: ALEXANDRIA, ARGENTINA, AUSTRALIA, AUSTRIA, BA |
| Recall initiated | 20120925 |
| Recall number | Z-0050-2013 |
| Category | device |
| State | MI |
| Country | United States |