← RecallGuard
HomeRecalls

The Anspach Effort, Inc. Recall: ANSPACH - Fluted Barrel; 8MM x 12.5MM REF L-8BA-G1 Cuttin

The Anspach Effort, Inc. Recall: ANSPACH - Fluted Barrel; 8MM x 12.5MM REF L-8BA-G1 Cuttin is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: The Anspach Effort, Inc.; Product: ANSPACH - Fluted Barrel; 8MM x 12.5MM REF L-8BA-G1 Cutting shaping bone including spine and cranium.; Reason for

Recalling firmThe Anspach Effort, Inc.
ProductANSPACH - Fluted Barrel; 8MM x 12.5MM REF L-8BA-G1 Cutting shaping bone including spine and cranium.
Reason for recallG1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.
Recall initiated20150413
Recall numberZ-2286-2015
Categorydevice
StateFL
CountryUnited States

🔍 Search all recalls →