The Metrix Company Recall: AUTOCOMP6 XPS High Speed Compounder REF 58810
The Metrix Company Recall: AUTOCOMP6 XPS High Speed Compounder REF 58810 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: The Metrix Company; Product: AUTOCOMP6 XPS High Speed Compounder REF 58810; Reason for recall: An issue was identified after a redesign of the auto compounding devic
| Recalling firm | The Metrix Company |
|---|---|
| Product | AUTOCOMP6 XPS High Speed Compounder REF 58810 |
| Reason for recall | An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Product distributed to medical facilities in the following US states: AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, and WA. Product was also distributed internationally to the following countries: Australia, Canada, Dominican Republic, Malaysia, Turkey, and United K |
| Recall initiated | 20160623 |
| Recall number | Z-2257-2016 |
| Category | device |
| State | IA |
| Country | United States |