Tosoh Bioscience Inc Recall: ST-AIA PACK Testosterone; Part Number: 025204 Assay, Reproduc
Tosoh Bioscience Inc Recall: ST-AIA PACK Testosterone; Part Number: 025204 Assay, Reproduc is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Tosoh Bioscience Inc; Product: ST-AIA PACK Testosterone; Part Number: 025204 Assay, Reproductive Hormone; Reason for recall: Asfotase Alfa (Strensi
| Recalling firm | Tosoh Bioscience Inc |
|---|---|
| Product | ST-AIA PACK Testosterone; Part Number: 025204 Assay, Reproductive Hormone |
| Reason for recall | Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica. |
| Recall initiated | 20180305 |
| Recall number | Z-2093-2018 |
| Category | device |
| State | OH |
| Country | United States |