Tosoh Bioscience Inc Recall: ST-AIA PACK Testosterone; Part Number: 025204 Assay, Reproduc

Tosoh Bioscience Inc Recall: ST-AIA PACK Testosterone; Part Number: 025204 Assay, Reproduc is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Tosoh Bioscience Inc; Product: ST-AIA PACK Testosterone; Part Number: 025204 Assay, Reproductive Hormone; Reason for recall: Asfotase Alfa (Strensi

Recalling firmTosoh Bioscience Inc
ProductST-AIA PACK Testosterone; Part Number: 025204 Assay, Reproductive Hormone
Reason for recallAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.
Recall initiated20180305
Recall numberZ-2093-2018
Categorydevice
StateOH
CountryUnited States