TP Orthodontics, Inc. Recall: Jasper Vektor, Class II Correction Appliance, Part No. 610-5 is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: TP Orthodontics, Inc.; Product: Jasper Vektor, Class II Correction Appliance, Part No. 610-527R used for correcting class 2 malocclusions by applyi
| Recalling firm | TP Orthodontics, Inc. |
|---|---|
| Product | Jasper Vektor, Class II Correction Appliance, Part No. 610-527R used for correcting class 2 malocclusions by applying light forces on the teeth using a tube, arch wire and stops. |
| Reason for recall | The firm has received 106 complaints of the device breaking in the patient's mouth. Device breakage may cause injury to the patient. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Distributed to accounts in 20 states: AL, CA, CO, FL, HI, ID, IL, MA, ME, MI, NJ, NV, NY, OH, OR, PA, SC, TX, UT, and VA. Foreign distribution to United Arab Emirates, Australia, Belgium, Brazil, Canada, Germany, Spain, Finland, France, Great Britain, Hungary, Ireland, Italy, Japan, Mexico, Netherla |
| Recall initiated | 20180501 |
| Recall number | Z-1788-2018 |
| Category | device |
| State | IN |
| Country | United States |