US Endoscopy Group Inc Recall: Endoscopic retrieval device. The disposable Raptor grasping
US Endoscopy Group Inc Recall: Endoscopic retrieval device. The disposable Raptor grasping is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: US Endoscopy Group Inc; Product: Endoscopic retrieval device. The disposable Raptor grasping device in intended to retrieve foreign bodies or resec
| Recalling firm | US Endoscopy Group Inc |
|---|---|
| Product | Endoscopic retrieval device. The disposable Raptor grasping device in intended to retrieve foreign bodies or resected tissue specimens in endoscopic procedures of the upper and lower gastrointestinal tract or biliary tract. |
| Reason for recall | When the outer catheter is either coiled or in a contorted configuration outside of the endoscope and the patient, and excessive force is applied to the handle, the jaws may become inoperable and may become stuck in the open condition. |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide Distribution -- USA (nationwide) including the states of AL, AR, AZ, CA, FL, GA, HI, IA, IL, KY, LA, MD, ME, MI, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, and WI; and to the following countries: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, |
| Recall initiated | 20111114 |
| Recall number | Z-0006-2013 |
| Category | device |
| State | OH |
| Country | United States |