Varian Medical Systems, Inc. Recall: VARISOURCE Breast Template System for high-dose rate

Varian Medical Systems, Inc. Recall: VARISOURCE Breast Template System for high-dose rate is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Varian Medical Systems, Inc.; Product: VARISOURCE Breast Template System for high-dose rate Brachytherapy. ***INSTRUCTIONS FOR USE*** Breast Bridge

Recalling firmVarian Medical Systems, Inc.
ProductVARISOURCE Breast Template System for high-dose rate Brachytherapy. ***INSTRUCTIONS FOR USE*** Breast Bridge & Template System AL13011000 ***Manufacturer: Varian Medical Systems, Inc., 3100 Hansen Way, Palo Alto, CA 94304 USA
Reason for recallThe recall was initiated after Varian Medical became aware the Lexan templates of the Breast Bridge Template System could become bent and turn opaque after steam sterilization.
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide distribution: US (nationwide) distribution in CA and in the country of India.
Recall initiated20130809
Recall numberZ-0065-2014
Categorydevice
StateVA
CountryUnited States