Zimmer Biomet, Inc. Recall: Zimmer M/DN¿ Intramedullary Fixation Humeral Guide Wire - Bull
Zimmer Biomet, Inc. Recall: Zimmer M/DN¿ Intramedullary Fixation Humeral Guide Wire - Bull is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Zimmer Biomet, Inc.; Product: Zimmer M/DN¿ Intramedullary Fixation Humeral Guide Wire - Bullet Tip, 2.4 mm Diameter, 70 cm Length Item Number: 00-2
| Recalling firm | Zimmer Biomet, Inc. |
|---|---|
| Product | Zimmer M/DN¿ Intramedullary Fixation Humeral Guide Wire - Bullet Tip, 2.4 mm Diameter, 70 cm Length Item Number: 00-2255-026-00 Product Usage: A single use surgical instrument used during orthopaedic surgery for implantation of a prosthesis. |
| Reason for recall | Various Trauma Guide Wires- creases in the sealing area on the end opposite of the product's chevron opening compromising product sterility |
| Classification | Class II (moderate — may cause temporary or reversible health consequences) |
| Status | Terminated |
| Distribution | Worldwide - US Nationwide Distribution Foreign: CANADA AUSTRALIA BRAZIL CHILE HONG KONG JAPAN MALAYSIA NETHERLANDS NICARAGUA SINGAPORE TAIWAN |
| Recall initiated | 20181022 |
| Recall number | Z-0532-2019 |
| Category | device |
| State | IN |
| Country | United States |