Zimmer Biomet, Inc. Recall: Zimmer ZMS¿ Intramedullary Fixation Smooth Guide Wire, 2.4 mm

Zimmer Biomet, Inc. Recall: Zimmer ZMS¿ Intramedullary Fixation Smooth Guide Wire, 2.4 mm is one of 9,000 recalls in the RecallGuard dataset. Key details — Recalling firm: Zimmer Biomet, Inc.; Product: Zimmer ZMS¿ Intramedullary Fixation Smooth Guide Wire, 2.4 mm Diameter, 100 cm Length Item Number: 47-2237-033-00 Prod

Recalling firmZimmer Biomet, Inc.
ProductZimmer ZMS¿ Intramedullary Fixation Smooth Guide Wire, 2.4 mm Diameter, 100 cm Length Item Number: 47-2237-033-00 Product Usage: A single use surgical instrument used during orthopaedic surgery for implantation of a prosthesis.
Reason for recallVarious Trauma Guide Wires- creases in the sealing area on the end opposite of the product's chevron opening compromising product sterility
ClassificationClass II (moderate — may cause temporary or reversible health consequences)
StatusTerminated
DistributionWorldwide - US Nationwide Distribution Foreign: CANADA AUSTRALIA BRAZIL CHILE HONG KONG JAPAN MALAYSIA NETHERLANDS NICARAGUA SINGAPORE TAIWAN
Recall initiated20181022
Recall numberZ-0533-2019
Categorydevice
StateIN
CountryUnited States